MODEL 3180/3190 - WAFER BONDERS
![]() |
|
| FOR
HIGH RELIABILITY WAFER BONDING Models 3180 and 3190 are programmable wafer bonders offered by SST International for high-reliability bonding of silicon, gallium-arsenide and glass wafers up to 6 inches (150 mm) in diameter. Both systems provide precise automatic control of heating and cooling at temperatures up to 500 °C (1000 °C optional). Bellows actuated clamping platens present uniform mechanical bonding forces up to 750 pounds (5,000 pounds optional) to the wafer pair. The 3190 also provides anodic (electrostatic) bonding capabilities with a user-selected high-voltage power supply and chamber feed-through. Vacuum levels down to 50 millitorr are provided in the 3180 by a two-stage mechanical pump. Vacuum levels below 1 x 10-6 torr are provided in the 3190 by a cryogenic vacuum pumping system combined with a dry mechanical forepump. Digital vacuum level gauging is provided to monitor and control vacuum levels. Machine control is provided by an embedded control system operating in a Microsoft Windows® environment. An unlimited number of process profiles can easily be created and stored in the controller. Run data is archived for quality control and off-line data analysis. Internet and intranet network connectivity is available as an option, permitting remote monitoring, troubleshooting and maintenance capabilities. |
|
TYPICAL APPLICATIONS: |
|
•
Silicon to Glass Wafer Bonding |
•
MEMS Wafer Bonding |
CONTROL SYSTEM Models 3180 and 3190 utilize distributed logic systems that link intelligent temperature and pressure controllers to a Microsoft Windows® based central control. All process parameters are controlled automatically with user-edited programs for each application. Multiple temperature ramp and soak cycles, vacuum and pressure cycling, mechanical wafer clamping force and anodic power supply settings are easily programmed as time-based events. Furnace operating characteristics are continuously monitored and the operator is alerted to any fault conditions. The operator interface features a color touch-screen display and PC keyboard. All process file results, including ultimate vacuum levels, are stored directly on the system hard drive. |
![]() Process Chamber |
|
![]() Color Touch-Screen Display |
SPECIFICATIONS*
3180 |
3190 |
|
| Vacuum Pump Type | Mechanical |
Cryogenic |
| Minimum Vacuum Level | 50
millitorr (.065 mbar) |
1
x 10-6 torr (1 x 10-6 mbar) |
| Operating Temperature Range | RT
to 500 °C (1000 °C option) |
|
| Thermal Work Zone | 6
in (150 mm) diameter |
|
| Adjustable Mechanical Clamping Force | 0
- 750 lb (0 - 3 kN) or 0 - 5,000 lb (0 - 20 kN) option |
|
| Maximum Chamber Gas Pressure Level | 50
psig (4.5 bar) |
15
psig (2 bar) |
| Process Gasses (three inputs) |
N2
required, (Ar, He, forming gas optional) @ 90 psig (7 bar) minimum pressure |
|
| Electrical Service | 208-240
volts, 60 amps, 60/50 Hz, 1 phase, 5 kilowatt average, 15 kilowatt peak |
|
| Cooling Water Required | 2
GPM (8 lpm) @ 20-25 °C, 2 kilowatt capacity minimum |
|
| Compressed Air Required | 90
psig (7 bar), ¼ in (6.5 mm) inside diameter line |
|
| Overall Size (W x D x H) |
58
x 54 x 53 in (147 x 137 x 135 cm) |
94
x 54 x 53 in (239 x 137 x 135 cm) |
| Helium Compressor Size (W x D x H) | not
applicable |
20
x 22 x 17 in (50 x 57 x 43 cm) |
| Total Weight | 1400
lb (650 kg) |
2400
lb (1100 kg) |
* Specifications subject to change